Specification name Specification contents
Bridge power supply 10 VDC±0.3 VDC within 30 mA (Changeable to 2.5 VDC) with remote sensing function
Input range F.S. setting is availableat the input range from -0.2 mV/V to 3.1 mV/V (When bridge power supply is DC5 V)
Zero adjustment range -1.9 mV/V ~ 1.9 mV/V
Non-linearity 0.01 %F.S.
Temperature coefficient Zero point ±0.2 μV/℃, Sensitivity ±0.001 5 %F.S./℃
(input conversion, F.S. setting at the input from 0.3 mV/V to 3.1 mV/V)
A/D sampling 100 times/s (Changeable to 4 times/s or 20 times/s)
Load display four systems of WS, DS, ADD and DIFF
Display range -19 999 ~ 99 999
Condition display HEALTHY, CHECK, DS DOWN, WS DOWN
Decimal point display Changeable to No display, 10¹, 10², 10³ or 10⁴
External control function ZERO, LOCK, DS DOWN, WS DOWN, DS & WS DOWN
Analog output(4ch) 0 VDC ~ 10 VDC, 0 VDC ~ 5 VDC, 0 VDC ~ 1 VDC, ±10 VDC, ±5 VDC, ±1 VDC load resistance 5 kΩor more, or 4 mADC ~ 20 mADC load resistane 510 Ωor less (changeable by the DIPswitch)
Operating temp./humidity range Temperature -10 ℃ ~ 50 ℃, Humidity 85 %RH or less (Non condensing)
Power supply 100 VAC ~ 240 VAC (85 VAC ~ 264 VAC) 50/60 Hz
Power consumption Approx. 25 VA
Dimension (W x H x D) 200 mm x 150 mm x 200 mm (excludes protruding parts)
Weight Approx. 4.4 kg (without options)