Move to the body of this page
Strain Amplifier

DC type dynamic strain amplifier DAS-406C

Features

・ DC ~ 200 kHz Characteristics in the frequency
・ The standard equivalent strain is set with the 4-digits digital switch
・ 3 1/2 digit digital monitor

Specification

Specification name Specification contents
Measuring point 1 point/1unit
Applicable gage resistance 60 Ω ~ 1kΩ
Gage factor 2
Bridge power supply 2 VDC (Changeable to 1 VDC, 4 VDC and 10 VDC )
with remote sensing function (50 mA or less)
Balancing adjustable range Resistance section Approx. ±1 % (Gage resistance  120 Ω Approx. ±5 000 μst)
Balancing adjustable method Auto balance, Auto balance time Approx.1 s,
Accuracy ±0.1 %F.S.
Sensitivity 0.5 VDC output at theinput of 100 μst (When bridge power supply is 2 VDC)
Output MONITOR ±10 V, ±5 mA (Load resistance2 kΩ or more)
OUTPUT ±10 V, ±30 mA (Load resistance 330 Ω or more)
Output resistance 10 Ω or less
Output capacitive load 0.1 μF or less
Non-linearity ±0.01 % F.S.
Standard equivalent strain(CAL) Set with digital switch, ±1 ~ ±9 999 μst, Accuracy ±0.2 %F.S. or less at each point
Sensitivity adjustablerange(ATT) 1, 1/2, 1/5, 1/10, 1/20, 1/50, 1/100, 0
Accuracy ±0.2 %F.S. at each point
GAIN volumu valiable range
1 to 1/2.5 times continuous vary(VAR)
Frequency response range DC to 200 kHz +1 dB, -3 dB (Filter at W/B)
Low-pass filter 10 Hz, 100 Hz, 1 kHz,10 kHz, W/B Third bessel type
S/N ratio 46 dBp-p(ATT=1) withW/B condition
55 dBp-p with 10 Hz to 10 kHz
Display Output voltage display 0.00to ±10.00
Digital display (Green LED)
Sampling Approx.4 times/s
0.00 or -0.00 flashes when output voltage is innon-linearity zone
Temperature efficient Zero shift ±1 μst/ ℃, Sensitivity variation ±0.01 %F.S./℃
Efficient due to time Zero shift ±5 μst/ 8h, Sensitivity variation ±0.05 %F.S./8 h
Efficient due to powersupply Zero shift ±1 μst/ variationof power supply voltage within ±10 %
Sensitivity valliation ±0.01 %F.S.
/variation of power supply voltage within ±10 %
Operating temp./humidity range Temperature :-10 ℃to 50 ℃, Humidity 85 %RHor less (Non condensing)
Power supply voltage 100 VAC±10 VAC, 50/60 Hz
Power consumption Approx. 10 VA(at AC100 V)
Outline dimensions(WxHxD) 42 mmx176 mmx284.4mm(excludes protruding parts)
Weight Approx. 1.5 kg

Download

Instruction Manual

Specification

Page Top