Introduction of the strain amplifier to amplify the signal from strain gage, transmit and indicate it, and the strain generator to calibrate the equipment connected with strain gage applied transducer, such as load cell, ressure sensor, etc.
Features
・Eqquipped the wire length correcting function and open circuit check function.
・High sensitivity and high gain
・Excellent visibility and operativeness
・Two power supply types
Alternative
A&D Company
Limited Industrial Measuring Instruments/Signal Conditioners/AC Strain Amplifier AS3000 Series/Strain amplifier AC bridge method
general-purpose
low bandwidth type AS3603
-
-
scroll
Specification name
Specification contents
Measuring point 1 point/1unit Applicable gage resistance 60 Ω ~ 1kΩ Gage factor 2 Bridge power supply 0.5 VACrms, 2 VACrms, 5 kHz sine wave,
Sync input/output signal 2.5 VACrmsBalancing adjustable range Deviation of resistive value :Approx. ±2 % (±10 000 μst)
Capacitive imbalance : Approx. 2 000 pF
Balancing adjusting Accuracy ±0.4 μst (RANGE=200, without FINE, BV=2 V)Maximum input range ±200 000 μst (RANGE=20k, FINE=x2.5, BV=0.5 V) Sensitivity ±10 V within ±200 μst Shift of measuring range 200, 500, 1k, 2k, 5k, 10k, 20k( μst, x2/BV in value), OFF
Fine adjustment Continuously changeable in FINE ~ RANGE
2 steps changing amount can be selected.Non-linearity ±0.01 % F.S. Standard equivalent strain(CAL) Set by digi-switch : ±1 to9 999 μst
Accuracy ±0.5 %rdg+0.5 μstFrequency response range DC to 2 kHz±10 % Hi-pass filter 0.5 Hz : 2-pole butterworth type (Filter descent response : -12 dB/oct) Low-pass filter 10 Hz, 30 Hz, 100 Hz, 300Hz, 500 Hz : 4-pole butterworth type (Filter descent response : -24 dB/oct) Temperature efficient Zero drift : Within ±0.1 μst/℃, Sensitivity : Within ±0.05 %F.S./℃ Efficient due to time Zero drift : Within ±0.5 μst /24h, Sensitivity : Within ±0.2 %F.S./24 h Noise 2.0 x10-6 in strain p-p RTI
(W/B, RANGE=200, Without FINE, BV=2 V, 120 Ω in bridge)
0.6 x 10-6 in strain p-p RTI
(DC ~ 100 Hz, RANGE=200, Without FINE, BV=2 V, 120 Ω in bridge)Output OUTPUT1 ±10 V, ±5 mA, Load resistance 2 kΩ
OUTPUT2 ±10 V, ±10 mA, Load resistance 1 kΩ
Output impedance 0.5 Ω or less, Capacitive load Operable up to 0.1 μFOutput adjustment OUTPUT2 ADJ (Can be independently varied continuously from 1 to 1/10) Operating temp./humidity range Temp. -10 ℃ to 50 ℃,
Humidity20 % to 85 %RH or less(Non condensing)Power supply voltage, Power consumption AC85 V to AC132 V/AC180V to AC264 V
(Internal switch must be changed) 9 VA or less
DC10 V to DC30 V, 6 VA or lessOutline dimensions(WxHxD) 49.5 mm x 143 mm x 253mm (excludes protruding parts) Weight Approx. 1.4 kg -
-
- Current output DSA-631-07
- Bridge Box Unit FA410-072
Custom
- ※Custom products are intended for mass production. Please consult us for small-lot orders.